EasyManua.ls Logo

Chroma 63600 Series - Communicating with Instruments

Default Icon
258 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Operation
5-11
STEP 2:
Input cmd and click OK to run the cmd program.
STEP 3:
A window of MS-DOS operation environment will open. Input “ping IP address” such as ping
10.1.9.20. If there is a response, it means the setting of local area network is done
successfully.
5.1.4.4 Communicating with Instruments
Users can use the application NI-MAX (Measurement & Automation Explorer) of National
Instruments to communicate the existing instruments or user developed application. To use
NI VISA, users need to open VISA Session Resource Name in the format of TCPIP0::<IP
address>::2101::SOCKET, for example, TCPIP0::10.1.7.100:: 2101::SOCKET. Otherwise,
specify the TCP/IP SOCKET PORT to 2101 if not using NI VISA.
Following is the example of using NI-MAX (Measurement & Automation Explorer) application.

Table of Contents

Related product manuals