5.3.2.10 MEASURE Subsystem
:MEASure
:CURRent
:VOLTage
:POWer
?
?
?
:ALLVoltage
:ALLCurrent
:ALLPower
?
?
?
<NR2>
<NR2>
<NR2>
<AARD>
<AADR>
<AARD>
:INPut
SP
?
LOAD / 0
UUT / 1
<CRD>
MEASure:ALLCurrent?
Type: Channel-Independent
Description: Returns the real time current measured at the input of all load
modules. The return value is 0 when the channel does not exist.
Query Syntax: MEASure:ALLCurrent?
Return Parameters:<aard>, [Unit = Ampere]
Query Example: MEAS:ALLC?
Return Example: 0, 0, 0, 0, 5.12, 0, 12, 0
MEASure:ALLPower?
Type: Channel-Independent
Description: Returns the real time power measured at the input of all load
modules. The return value is 0 when the channel does not exist.
Query Syntax: MEASure:ALLPower?
Return Parameters:<aard>, [Unit = Watt]
Query Example: MEAS:ALLP?
Return Example: 0, 0, 0, 0, 5.12, 0, 12, 0