Programmable DC Electronic Load 63600 Series Operation & Programming Manual
5-90
5.3.2.20 VOLTAGE Subsystem
:STATic
:L2
:L1
SP
<NRf+>
?
SP
MAX
MIN
<NR2>
Suffix
<NR2>
<NR2>
:VOLTage
:ILIMit
:TYPE
:RESponse
SP
<NRf+>
?
SP
MAX
MIN
<NR2>
Suffix
<NR2>
<NR2>
SP
CURR/PSU / 0
VOLT/PSU / 1
?
SP
FAST / 0
NORMAL / 1
?
SLOW / 2
<CRD>
<CRD>
VOLTage:STAT:ILIMit
Type: Channel-Specific
Description: Set the current limit for constant voltage mode.
Setting Syntax: VOLTage:STATic:ILIMit<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example:
VOLT:STAT:ILIM 3 Set the current limit to 3A in
constant voltage mode.
VOLT:STAT:ILIM MAX Set the current limit to the
maximum value in constant
voltage mode.
VOLT:STAT:ILIM MIN Set the current limit to the
minimum value in constant
voltage mode.
Query Syntax: VOLTage:STATic:ILIMit?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Ampere]
Query Example: VOLT:STAT:ILIM?
VOLT:STAT:ILIM? MAX
VOLT:STAT:ILIM? MIN
VOLTage:STATic:L1
Type: Channel-Specific
Description: Set the static load voltage in constant voltage mode.
Setting Syntax: VOLTage:STATic:L1<space><NRf+>[suffix]