Page #195 background imageLoading...
Page #195 background image
Remote Operation
5-63
FETCh:ALLVoltage:MAX?
Type: Channel-Independent
Description: Returns the maximum voltage measured at the input of all load
modules. The return value is 0 when the channel does not exist.
Query Syntax: FETCh:ALLVoltage:MAX?
Return Parameters:<aard>, [Unit = Voltage]
Query Example: FETC:ALLV:MAX?
Return Example: 0, 0, 0, 0, 0, 0, 5.12, 0, 12, 0
FETCh:ALLVoltage:MIN?
Type: Channel-Independent
Description: Returns the minimum voltage measured at the input of all load
modules. The return value is 0 when the channel does not exist.
Query Syntax: FETCh:ALLVoltage:MIN?
Return Parameters:<aard>, [Unit = Voltage]
Query Example: FETC:ALLV:MIN?
Return Example: 0, 0, 0, 0, 0, 0, 5.12, 0, 12,0
5.3.2.8 IMPEDANCE Subsystem
:
STATic
:RL
:RS
:LS
SP
<NRf+>
?
SP
MAX
MIN
SP
<NRf
+>
?
SP
MAX
MIN
<NR2>
Suffix
<NR2
>
<NR2>
<NR2>
<NR2>
<NR2>
:
IMPedance
:CL
Suffix
SP
<
NRf+>
?
SP
MAX
MIN
<NR
2>
Suffix
<
NR2>
<NR2>
SP
<NRf+>
?
SP
MAX
MIN
<NR
2>
<NR2>
<NR2>
Suffix
IMPedance:STATic:CL
Type: Channel-Specific
Description: Set the equivalent parallel load capacitance for constant impedance
mode.
Setting Syntax: IMPedance:STATic:CL<space><NRf+>[suffix]
Setting Parameters:<NRf+>, 30μF ~ 50,000μF, Resolution = 1uF, Unit = Farad
Setting Example: IMP:STAT:CL 0.02 Set capacitance = 20mF.
IMP:STAT:CL 100μF Set capacitance = 100μF.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Chroma 63600 Series and is the answer not in the manual?

Chroma 63600 Series Specifications

General IconGeneral
Series63600
Operating Temperature0°C to 40°C
ModesCC, CV
InterfaceGPIB, USB
ProtectionOCP, OTP, OVP
Voltage Resolution1mV
CoolingForced Air