Programmable DC Electronic Load 63600 Series Operation & Programming Manual
IMP:STAT:CL MAX Set capacitance = maximum value.
IMP:STAT:CL MIN Set capacitance = minimum value.
Query Syntax: CURRent:STATic:CL?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Farad]
Query Example: CURR:STAT:CL?
CURR:STAT:CL? MAX
CURR:STAT:CL? MIN
IMPedance:STATic:LS
Type: Channel-Specific
Description: Set the equivalent series inductance for constant impedance mode.
Setting Syntax: IMPedance:STATic:LS<space><NRf+>[suffix]
Setting Parameters:<NRf+>, 0 ~ 20.0μH, Resolution = 0.1μH, Unit = Henry
Setting Example: IMP:STAT:LS 0.00002 Set inductance = 20μH.
IMP:STAT:LS 1μH Set inductance = 1μH.
IMP:STAT:LS MAX Set inductance = maximum value.
IMP:STAT:LS MIN Set inductance = minimum value.
Query Syntax: CURRent:STATic:LS?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Henry]
Query Example: CURR:STAT:LS?
CURR:STAT:LS? MAX
CURR:STAT:LS? MIN
IMPedance:STATic:RS
Type: Channel-Specific
Description: Set the equivalent series resistance for constant impedance mode.
Setting Syntax: IMPedance:STATic:RS<space><NRf+>[suffix]
Setting Parameters:<NRf+>, 0.03Ω ~ 20.00Ω, Resolution = 0.01Ω, Unit = Ohm
Setting Example: IMP:STAT:RS 20 Set resistance = 20Ω.
IMP:STAT:RS 10 OHM Set resistance = 10Ω.
IMP:STAT:RS MAX Set resistance = maximum value.
IMP:STAT:RS MIN Set resistance = minimum value.
Query Syntax: IMPedance:STATic:RS?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = OHM]
Query Example: IMP:STAT:RS?
IMP:STAT:RS? MAX
IMP:STAT:RS? MIN
IMPedance:STATic:RL
Type: Channel-Specific
Description: Set the equivalent parallel load resistance for constant impedance
mode.
Setting Syntax: IMPedance:STATic:RL<space><NRf+>[suffix]
Setting Parameters:For valid value range refer to respective specification.
Setting Example: IMP:STAT:RL 20 Set resistance = 20Ω.
IMP:STAT:RL 10 OHM Set resistance = 10Ω.
IMP:STAT:RL MAX Set resistance = maximum value.
IMP:STAT:RL MIN Set resistance = minimum value.
Query Syntax: IMPedance:STATic:RL?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = OHM]
Query Example: IMP:STAT:RL?
IMP:STAT:RL? MAX
IMP:STAT:RL? MIN