Set Up
In order to use the xHV Power Supply, the PC running your CH2 tester must have 2 USB
ports. The xHV Power Supply requires the current version of the Easy-Wire software (version
2009.2.0 or higher) and a driver (see below). Prepare all cords to be assembled.
The version of your Easy-Wire software is
found on the Easy-Wire main menu.
If you need to install the current version
of the Easy-Wire software, go to
www.cirris.com. A password will be
needed to access the software. To
obtain your password, call Cirris
customer support at 1-800-441-9910.
Installing the Driver
Ensure that Easy-Wire version 2009.2.0 or higher is installed on your PC. Then navigate to the
file path C:\Program Files\Cirris\easywire\xHVAC Supply, double-click USBXpressInstaller.exe,
and follow the installation steps provided. This installation may take a few minutes.
The electrical assembly of the CH2 Base
should already be completed.
Setting Up the xHV Power Supply:
1. Place the CH2 Base on top of the CH2
xHV Power Supply aligning the back
and side panels of both units as
shown.
xHV power
supply
(rear view)