- 101 -
-
-
-
Message Cause Alarm Level
T 1 TOO HIGH
Measuring value of channel 1 is
above upper alarm limit.
User-selectable
TI TOO LOW
Measuring value of channel 1 is
below lower alarm limit.
User-selectable
T2 TOO HIGH
Measuring value of channel 2 is
above upper alarm limit.
User-selectable
T2 TOO LOW
Measuring value of channel 2 is
below lower alarm limit.
User-selectable
TD TOO HIGH
Difference between two channels is
larger than upper limit.
User-selectable
Technical alarms:
Alarm Message Cause
Alarm
Level
Remedy
T 1 SENSOR OFF
Temperature
cable of channel
1 may be
disconnected
from the monitor.
LOW
Make sure that the cable is
properly connected.
T2 SENSOR OFF
Temperature cable
of channel 2 may be
disconnected from
the monitor.
LOW
Make sure that the cable is
properly connected.
T1 ALM LMT ERR
Functional safety
failure
HIGH
Stop using alarming function
of TEMP module, notify
biomedical engineer or Our
service staff.
T 2 ALM LMT ERR
Functional safety
failure
HIGH
Stop using alarming function
of TEMP module, notify
biomedical engineer or Our
service staff.
TD ALM LMT ERR
Functional safety
failure
HIGH
Stop using alarming function
of TEMP module, notify
biomedical engineer or Our
service staff.
Prompt message:
Message Cause Alarm Level
T1 EXCEED
Measuring value of channel 1 is
beyond measuring range.
HIGH
T 2 EXCEED
Measuring value of channel 2 is
beyond measuring range.
HIGH
14.4
14.4
14.4
14.4 Care
Care
Care
Care and
and
and
and Cleaning
Cleaning
Cleaning
Cleaning
Warning
Warning
Warning
Warning
Before
Before
Before
Before cleaning
cleaning
cleaning
cleaning the
the
the
the monitor
monitor
monitor
monitor or
or
or
or the
the
the
the probe,
probe,
probe,
probe, make
make
make
make sure
sure
sure
sure that
that
that
that the
the
the
the equipment
equipment
equipment
equipment is
is
is
is switched
switched
switched
switched off
off
off
off and
and
and
and
disconnected
disconnected
disconnected
disconnected from
from
from
from the
the
the
the power
power
power
power line.
line.
line.
line.