- 68 -
-
-
-
Physiological
Physiological
Physiological
Physiological alarms:
alarms:
alarms:
alarms:
Message Cause Alarm level
ECG LOST No ECG signal of the patient is detected. HIGH
HR TOO HIGH HR measuring value is above the upper alarm limit User-selectable
HR TOO LOW HR measuring value is below the lower alarm limit User-selectable
Technical
Technical
Technical
Technical alarms:
alarms:
alarms:
alarms:
Message Cause
Alarm
level
Remedy
ECG LEAD OFF
ECG electrodes fall off
the skin or ECG cables fall off
the monitor.
LOW
Make sure that all
electrodes, leads and
patient cables are
properly connected.
ECG V LEAD OFF or
ECG C LEAD OFF
ECG LL LEAD OFF or
ECG F LEAD OFF
ECG LA LEAD OFF or
ECG L LEAD OFF
ECG RA LEAD OFF or
ECG R LEAD OFF
ECG INIT ERR
ECG module failure HIGH
Stop using
measuring function
provided by ECG
module, notifies
biomedical engineer
or Our service staff.
ECG INIT ERR 1
ECG INIT ERR 2
ECG INIT ERR 3
ECG INIT ERR4
ECG INIT ERR5
ECG INIT ERR 6
ECG INIT ERR7
ECG INIT ERR8
ECG COMM STOP
Occasional communication
failure
HIGH
If failure persists,
notify biomedical
engineer or Our
service staff.
ECG COMM ERR
Occasional communication
failure
HIGH
If failure persists,
notify biomedical
engineer or Our
service staff.
HR ALM LMT ERR Functional safety failure HIGH
Stop using HR alarm
function, notify
biomedical engineer
or Our service staff.

Other manuals for Contec CMS8000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Contec CMS8000 and is the answer not in the manual?

Contec CMS8000 Specifications

General IconGeneral
CategoryMedical Equipment
Device NameContec CMS8000
TypePatient Monitor
ECG Lead Mode3-lead or 5-lead
Resolution800 x 600
Power SupplyAC 100-240V, 50/60Hz
BatteryRechargeable lithium battery
Operating Temperature5°C to 40°C
Storage Temperature-20°C to 55°C
DisplayTFT LCD
Screen Size12.1 inches
ParametersECG, RESP, NIBP, SpO2, TEMP, PR
InterfaceEthernet, Nurse call
Humidity15% to 95% non-condensing
Battery LifeApproximately 4 hours