- 86 -
-
-
-
occurring during SpO
2
measurement.
Physiological alarm:
Message Cause Alarm Level
S P O
2
TOO HIGH
SpO
2
measuring value is above upper alarm limit.
User-selectable
S p O
2
TOO LOW
SpO
2
measuring value is below lower alarm limit.
User-selectable
PR TOO HIGH
PR measuring value is above upper alarm limit.
User-selectable
PR TOO LOW PR measuring value is below lower alarm limit. User-selectable
Technical alarms:
Message Cause
Alarm
Level
Remedy
SPO
2
SENSOR OFF
SpO
2
sensor may
be disconnected
from the patient or
the monitor.
LOW
Make sure that the
monitor and the patient
are in correct connection
with the cables.
SPO
2
INIT ERR
SpO
2
module
failure
HIGH
Stop using the
measuring function of
SpO
2
module, notify
biomedical engineer or
o ur service staff.
SPO
2
INIT ERR 1
SPO
2
INIT ERR 2
SPO
2
INIT ERR 3
SPO
2
INIT ERR 4
SPO
2
INIT ERR 5
SPO
2
INIT ERR 6
SPO
2
INIT ERR 7
SPO
2
INIT ERR 8
SPO
2
COMM STOP
SpO
2
module
failure or
communication
error
HIGH
Stop using the
measuring function of
SpO
2
module, notify
biomedical engineer or
o ur service staff.
SPO
2
COMM ERR
SpO
2
module
failure or
communication
error
HIGH
Stop using the
measuring function of
SpO
2
module, notify
biomedical engineer or
our service staff.
SPO
2
ALM L M T ERR
Functional safety
failure
HIGH
Stop using the
measuring function of
SpO
2
module, notify
biomedical engineer or
our service staff.
PR ALM LMT ERR
Functional safety
failure
HIGH
Stop using the
measuring function of
SpO
2
module, notify
biomedical engineer or
our service staff.

Other manuals for Contec CMS8000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Contec CMS8000 and is the answer not in the manual?

Contec CMS8000 Specifications

General IconGeneral
CategoryMedical Equipment
Device NameContec CMS8000
TypePatient Monitor
ECG Lead Mode3-lead or 5-lead
Resolution800 x 600
Power SupplyAC 100-240V, 50/60Hz
BatteryRechargeable lithium battery
Operating Temperature5°C to 40°C
Storage Temperature-20°C to 55°C
DisplayTFT LCD
Screen Size12.1 inches
ParametersECG, RESP, NIBP, SpO2, TEMP, PR
InterfaceEthernet, Nurse call
Humidity15% to 95% non-condensing
Battery LifeApproximately 4 hours