EasyManuals Logo

Contec CMS8000 Manual

Contec CMS8000
153 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #88 background imageLoading...
Page #88 background image
-
-
-
- 85 -
-
-
-
SpO
SpO
SpO
SpO
2
2
2
2
alarm
alarm
alarm
alarm setting
setting
setting
setting
ALM: pick "ON", the system will give alarm prompt and store alarm information when SpO
2
alarm
occurs; pick "OFF", the system will not give alarm and instead display a beside SpO
2
.
ALM REC: pick "ON", the system will command the recorder to output alarm information when SpO
2
alarm occurs.
ALM LEV: used to set up alarm level , selectable from HIGH, MED and LOW . HIGH represents the
most serious case.
SPO
2
ALM HI and SPO
2
ALM LO: SpO
2
alarm is activated when the result exceeds set SPO
2
ALM
HI value or falls below SPO
2
ALM LO value.
PR ALM HI and PR ALM LO: PR alarm is activated when the pulse rate exceeds set PR ALM HI
value or falls below PR ALM LO value.
SpO
2
and PR alarm limits:
Max. Upper Limit M in . Lower Limit Step
SpO
2
100 0 1
PR 25 0 0 1
The default SpO
2
and PR alarm limits:
Parameters
Max. Upper Limit M in . Lower Limit
SpO
2
Adult 100 90
Pediatric 100 90
Neonatal 95 85
PR
Adult 120 50
Pediatric 160 75
Neonatal 200 100
SWEEP
Available options are 12.5mm/s, 25.0 mm/s.
AVG T IME
4 s , 8 s , 16s represent times that SpO
2
average value is counted.
DEFAULT:
P ick this item to access the SPO
2
DEFAULT CONFIG dialog box, in which you can select
FACTORY DEFAULT CONFIG or USER DEFAULT CONFIG. After selecting one item and exiting
the dialog box, the system will pop up the dialog box asking for the your confirmation.
12.6
12.6
12.6
12.6 Alarm
Alarm
Alarm
Alarm Description
Description
Description
Description and
and
and
and Prompt
Prompt
Prompt
Prompt
SpO
SpO
SpO
SpO
2
2
2
2
Alarm
Alarm
Alarm
Alarm Message
Message
Message
Message
When the alarm switches are set to ON in relevant menus, the physiological alarms caused by the
parameter exceeding the alarm limit may possibly trigger the recorder to automatically output alarming
parameter value and corresponding waveforms.
Tables below describe the possible physiological alarms, technical alarms and prompt messages

Other manuals for Contec CMS8000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Contec CMS8000 and is the answer not in the manual?

Contec CMS8000 Specifications

General IconGeneral
CategoryMedical Equipment
Device NameContec CMS8000
TypePatient Monitor
ECG Lead Mode3-lead or 5-lead
Resolution800 x 600
Power SupplyAC 100-240V, 50/60Hz
BatteryRechargeable lithium battery
Operating Temperature5°C to 40°C
Storage Temperature-20°C to 55°C
DisplayTFT LCD
Screen Size12.1 inches
ParametersECG, RESP, NIBP, SpO2, TEMP, PR
InterfaceEthernet, Nurse call
Humidity15% to 95% non-condensing
Battery LifeApproximately 4 hours

Related product manuals