- 96 -
-
-
-
hoses are tangled. engineer or Our service staff.
WEAK SIGNAL
Cuff is too loose or
patient pulse is too
weak.
LOW
Use other method to measure blood
pressure.
RANGE
EXCEEDED
Measuring range
exceeds the specified
upper limit.
HIGH
Reset NIBP module, if failure
persists, stop using measuring
function of NIBP module, notify
biomedical engineer or Our service
staff.
EXCESSIVE
MOTION
Affected by arm
motion, signal noise is
too large or pulse rate
is not regular.
LOW
Make sure that the patient under
monitoring is motionless.
OVER PRESSURE
Pressure has
exceeded the
specified upper safety
limit.
HIGH
Measure again, if failure persists,
stop using measuring function of
NIBP module and notify biomedical
engineer or Our service staff.
SIGNAL
STURATED
Excessive motion LOW Stop the patient from moving.
PNEUMATIC LEAK
During pneumatic test,
leak is detected.
LOW
Check and replace the leaking parts,
if required, notify biomedical engineer
or Our service staff.
NIBP SYSTEM
FAILURE
Operation of blood
pressure pump
system is failed.
HIGH
Stop using measuring function of
NIBP module, notify biomedical
engineer or Our service staff.
CUFF TYPE ERR
Cuff type does not
comply with the
patient type.
LOW Select appropriate cuff type
NIBP TIME OUT
Measuring time has
exceeded 120
seconds (adult) or 90
seconds (neonatal).
HIGH
Measure again or use other
measuring method.
NIBP ILLEGALLY
RESET
Abnormal module
reset
HIGH Reset again
MEASURE FAIL
Problem happens
when measuring the
curve. The system
cannot perform
measurement,
analysis or
calculation.
HIGH
Check the cuff. Make sure that the
patient under monitoring is
motionless. Measure again.
Prompt message: (display in the prompt area below NIBP value)
Message Cause Alarm Level
Manual m easure During manual measuring mode.
No alarm
C ont measuring During continuous measuring mode.
A uto m easuring During automatic measuring mode.
Please start
After selecting interval between
measurements in MENU
Measurement over
Press START/STOP key during measuring
to stop measurement
Calibrating During calibrating
Calibration over Calibration over
Pneum testing During pneumatic test

Other manuals for Contec CMS8000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Contec CMS8000 and is the answer not in the manual?

Contec CMS8000 Specifications

General IconGeneral
CategoryMedical Equipment
Device NameContec CMS8000
TypePatient Monitor
ECG Lead Mode3-lead or 5-lead
Resolution800 x 600
Power SupplyAC 100-240V, 50/60Hz
BatteryRechargeable lithium battery
Operating Temperature5°C to 40°C
Storage Temperature-20°C to 55°C
DisplayTFT LCD
Screen Size12.1 inches
ParametersECG, RESP, NIBP, SpO2, TEMP, PR
InterfaceEthernet, Nurse call
Humidity15% to 95% non-condensing
Battery LifeApproximately 4 hours