- 95 -
-
-
-
13.4
13.4
13.4
13.4 NIBP
NIBP
NIBP
NIBP Alarm
Alarm
Alarm
Alarm Message
Message
Message
Message
Among physiological alarms, those belonging to the type that the parameter has exceeded the limits
may activate the recorder to automatically output the parameters and related measured waveforms
when the alarms occur on the condition that the alarm record switch in the related menu is On.
Tables below describe the possible physiological alarms, technical alarms and prompt messages
occurring during NIBP measurement.
Physiological alarms:
Message Cause Alarm Level
NS TOO HIGH
NIBP SYS measuring value is
above upper alarm limit.
User-selectable
NS TOO LOW
NIBP SYS measuring value is
below lower alarm limit.
User-selectable
ND TOO HIGH
NIBP DIA measuring value is
above upper alarm limit.
User-selectable
ND TOO LOW
NIBP DIA measuring value is
below lower alarm limit.
User-selectable
NM TOO HIGN
NIBP MAP measuring value
is above upper alarm limit.
User-selectable
NM TOO LOW
NIBP MAP measuring value
is below lower alarm limit.
User-selectable
Technical alarms 1: (display in information area)
Message Cause
Alarm
Level
Remedy
NS ALM LMT
ERR
Functional safety failure HIGH
Stop using alarming functions of
NIBP module and notify biomedical
engineer or Our service staff.
N M ALM LMT
ERR
Functional safety failure HIGH
Stop using alarming functions of
NIBP module and notify biomedical
engineer or Our service staff.
N D ALM LMT
ERR
Functional safety failure HIGH
Stop using alarming functions of
NIBP module and notify biomedical
engineer or Our service staff.
Technical alarms 2: (display in the area below the NIBP value)
Message Cause
Alarm
Level
Remedy
NIBP SELF TEST
ERR
Sensor or other
hardware of NIBP
module is incorrect.
HIGH
Stop using measuring function of
NIBP module, notify biomedical
engineer or Our service staff.
NIBP COMM ERR
Communication with
NIBP module is failed.
HIGH
If failure persists, stop using
measuring function of NIBP module,
notify biomedical engineer or Our
service staff.
LOOSE CUFF
Cuff is no properly
wrapped or no cuff
exists.
LOW Properly wrap the cuff
AIR LEAK
Cuff, hose or
connector is
damaged.
LOW
Check and replace the leaking parts,
if required, notify biomedical engineer
or Our service staff.
AIR PRESSURE
ERROR
Stable pressure value
is not available. e.g.
LOW
Check if the hoses are tangled, if
failure persists, notify biomedical