AutoLab TF Dispenser Technical Manual – Revision 5
AutoLab TF dispensers Technical Manual TFTMTF / Rev 5.0 / 24
th
May 2006 Page 26 of 150
1-6-2 AutoLab TF 88/128/168 System
1-6-2-1 Auxiliary/Water Dispensing
Auxiliary bottle
There are 7 bottles of auxiliary solution, and each one is 1000 ml.
Agitate device
Containing agitation mechanism for all auxiliary bottles, the power on/off and speed is
controlled by manual.
Water tank
This water tank with level sensor and automatic refill, and the size is 7 litres.
Auxiliary bottle
There are 7 auxiliary bottles for refill, and each capacity is 5 litres.
Auxiliary refill valve
The auxiliary auto refill is controlled by magnetic valve.
Air regulator
This device is used to regulate air pressure, and output to all cylinders and the refill air
regulator. The air pressure should not below 4.5~6.0 kg/cm
2
.
Refill air regulator
This device is used to regulate the air pressure for air gun and auxiliary refill system.
The air pressure should be reach 1 kg/cm
2
.
Water pressure regulator
This device is used to regulate the water inlet pressure for water auto refill valve. The
water pressure should be adjusted below 1 kg/cm
2
.
Water auto refill valve
This device is used for refilling water.
Air gun
This device is used for priming the tube of auxiliary.
Auxiliary dispense cylinder control valve
This device is used to control the cylinder that moves auxiliary dispensing head to front,
middle and back position.
Tray push in cylinder control valve
This device is used to control the cylinder that moves the tray push bar to auxiliary,
balance and back position.
Auxiliary calibrate cylinder control valve
This device is used to control the cylinder that moves auxiliary dispensing head to back
and calibrate position.
Tray up/down cylinder control valve
This device is used to control cylinder that move the tray to up and down position.