AutoLab TF Dispenser Technical Manual – Revision 5
AutoLab TF dispensers Technical Manual TFTMTF / Rev 5.0 / 24
th
May 2006 Page 88 of 150
2-4 Balance
2-4-1 The Parameter Setup Of Balance
Mettler Toledo PG6002S
The detailed description of the menu options is given as below:
- To entry the setup menu, hold down the “MENU” key about 3 seconds when
the power is on.
- Press “MENU” to switch the menu item one by one.
- Press ”FT” key to change the setting and hold down “MENU” key to save the
setting.
- Each time you can only change one setting, and repeat the above step to
change the others.
PS:Press “C” key exit and the setting don’t save.
The settings of scale should be same as below:
NO. FUNCTION EXPLAINATION AutoLab TF
1
RESET
Call-up the factory setting.
2
CAL OFF
Need engineer to enable calibration only.
3
INFO OFF
Don’t show the calibration info.
4
F none
No available in weighing operation at a keystroke.
5
~ 3
The ambient condition is unstable.
6
、 3
The weighing process is absolute weighing.
7
GOOD
The repeatability is good.
8
Unit 1 g
The 1
st
weighing unit is gram.
9
Unit 2 g
The 2
nd
weighing unit is gram.
10
AZ. OFF
The automatic zero correction is off.
11
A.OFF
The automatic shutdown is off.
12
qu.START
Start without display test.
13
ON
Display the icons.
14
HOST
Attachment to a host.
15
S.CONT
Data transfer mode is continuation.
16
S.SICS
Data transfer format is MT-SICS.
17
bd 9600
Data transfer speed is 9600bps.
18
8b-no
Character format is 8bit no parity.
19
HS OFF
Handshake is off.
20
Secure 2
Lock control panel
Note that it is recommended by Mettler that scales are left on to warm up for a period of
time before using them. The time recommended is as follows:
Mettler PG4002-S and PG-6002-S scales should be left switched on for 30 minutes for stability
warm up prior to using them.
Mettler PR5003-DR and PR-2003-DR scales should be left switched on for 60 minutes for
stability warm up prior to using them.