AutoLab TF Dispenser Technical Manual – Revision 5
AutoLab TF dispensers Technical Manual TFTMTF / Rev 5.0 / 24
th
May 2006 Page 94 of 150
2-4-2 Balance calibration
Mettler Scale
Internal Calibration
STEP1:Push the button on the plate for 3 seconds for automatic rectification.
STEP2: on the screen means rectification done.
External Calibration
(for 3, 4, 5, and 6 ㎏ scale)
STEP1:Push the button on the plate for 3 seconds, on the
screen.
STEP2:Push again for one time, on the screen.
STEP3:Push the button on the plate , and then on the screen.
STEP4:Push the button on the plate for 3 seconds, on the screen,
then back to main screen.
STEP5:Push the button on the plate for 3 seconds, on the
screen.
STEP6:Push the button on the plate for rectify the weight to 3000g.
STEP7:Push the button on the plate for 3 seconds, on the screen.
STEP8:Put calibrate weight 3 ㎏ for rectification, then on the screen flashing.
STEP9:Take down calibrate weight, on the screen for rectification done.
For changing to Inside rectification, Please have change to saved.
1/10d
Cal
Cal done
Menu Reset
Menu Fact
T
F
Vari Cal
Menu
Stored
1/10d
Cal
Cal 5000g
T
F
Menu
3000g
0.00g
Cal done
Vari Cal
Fact