Precisa XT6200C
Internal Calibration
STEP 1Switch to “BALANCING” with the change key.
STEP 2Press T until “CALIBRATION” is appears.
STEP 3The calibration is finished after a certain period of time.
External Calibration
STEP 1Switch to “BALANCING” with the change key.
STEP 2Press T until “CALIBRATION” is appears.
STEP 3The balance carries out a Zero measurement (0.000g is shown flashing).
STEP 4After the zero measurement the display flashes with the recommended calibration
weight.
STEP 5Place the calibration weight on the pan.
STEP 6The display continues to flash.
STEP 7Calibration is complete when the display stops flashing.