ES601Plus OPERATING MANUAL
The options available on the SAVE TEST RECORD menu are:
DEVICE EDIT:
This selection opens the DEVICE INFO menu as described earlier in
this chapter.
ERASE MEMORY:
Opens the Erase Memory menu. The user is provided with options for
erasing the last (i.e. most recent) Test Record stored in flash memory, or
all Records in flash memory.
DISCARD TEST
DATA:
Discards all device information and measurement data for the test in
progress and returns to the MANUAL MODE menu. The Clip Record
is erased by pressing this button.
KEEP TEST DATA:
Returns the display to the MANUAL MODE menu, and keeps the
measurement data and device information in the Clip Record.
SAVE TEST
RECORD:
Saves the test data to a Test Record in flash memory, and returns the
display to the MANUAL MODE menu. Make, model, location, facility
and Tech Code fields of the device information are retained, preparing
ES601Plus to test another similar device. All other device information
and data in the Clip Record are erased.
SAVE TEST AND
ADD MODULE:
Saves the test data according to the description above for SAVE TEST
RECORD, and creates a link to the next record to be saved through
the Parent ID field of the Test Record.
A practical application of this feature is to link results for a modular
patient monitor under the control number of the monitor's mainframe,
but where each plug-in module of the monitor could be independently
identified by its own control number or other ID in an equipment
management system.
The Parent ID is utilized by the Test Report Generator of the ES601pc
software package, which can use the Parent ID field within each Test
Record to generate a single integrated report for the mainframe and all
modules comprising a modular multi-parameter patient monitor. The
SAVE TEST AND ADD MODULE button will continue to link Test
Records for the modules via the Parent ID, until the Test Record for
the final module of the patient monitor is stored by pressing the SAVE
TEST RECORD button. This will break the chain of linked Records
by clearing the Parent ID field.
Manual Testing/Chapter 5 # Page 81