ES601Plus OPERATING MANUAL
APPENDIX A. Safety & ECG AutoSequences
The following are the default Safety AutoSequences of ES601Plus. ES601Plus is supplied with these
AutoSequences pre-loaded in the flash memory. The default AutoSequences may be modified using
the ES601pc software, supplied on CDROM with ES601Plus.
AutoSequences in the following list are identified as "SAFETY AUTOSEQUENCE #NN
(SAFNN)." (SAFNN) refers to the command assigned by Datrend to the AutoSequence, to allow
the AutoSequence to be invoked from a Test Procedure (see Au to m ate d T e stin g Prin c ip le s ,
Chapter 6; and Te s t Pro c e d u re Ed ito r, Chapter 8).
SAFETY AUTOSEQUENCE 1 (SAF01)
IEC601 SAFETY TEST
CHASSIS LKG & GND
Standard: IEC60601
Limits Used: IEC60601 Limits
Equipment Class: Class I
Line Voltage: Measure
Load Current: Skip
Insulation Resistance - Line 1 and 2 to Ground: Skip
Insulation Resistance - Applied Parts to Ground: Skip
Protective Earth: Measured At 1 Amp
Leakage Current: Measure
Device Power: On
Polarity: Normal/Reverse
Line 2: Normal/Open
Ground: Normal/Open
Test Enclosure and Earth Plus:
Applied Part Groups
None
Appendix A # Page 177