ES601Plus OPERATING MANUAL
The ESCAPE key may be pressed on any submenu to return to the main ECG menu. On the main
ECG menu, pressing ESCAPE returns the display to the MANUAL MODE menu.
When manually operating the ECG Simulator via the prior menu system, it is not possible to save
"ECG test results" in the Clip Record. ECG test results can be obtained and saved to the Clip
Record by running an ECG AutoSequence as described in the ECG AutoSequence section of
Chapter 6.
5.16 Technician Timer
Some institutions require monitoring and recording of the time spent performing certain types of
work. This requirement is satisfied in ES601Plus by an elapsed time counter, which is started and
stopped in response to activity of the analyzer.
When ES601Plus is powered up, the "tech time" counter is initially reset and then stopped.
When any measurement task is selected, the timer is started if it is not currently running. The timer
counts upward in elapsed seconds and has sufficient range to clock up to 18 continuous hours of
work. The timer continues to run until a Test Record is saved by pressing SAVE TEST. The timer
value is then saved in the record along with the measurement data.
Operation of the Tech Time counter is similar for all tests performed by ES601Plus, including those
tests selected from the AUTOSEQUENCE MODE menu.
Manual Testing/Chapter 5 # Page 97