ES601Plus OPERATING MANUAL
SAFETY AUTOSEQUENCE 15 (SAF15)
AAMI SAFETY TEST
3 LEADS (ECG)
Standard: AAMI ES1
Limits Used: AAMI ES1 Limits
Equipment Class: Class I
Line Voltage: Measure
Load Current: Measure
Insulation Resistance - Line 1 and 2 to Ground: Measure
Insulation Resistance - Applied Parts to Ground: Measure
Protective Earth: Measured At 1 Amp
Leakage Current: Measure
Device Power: On
Polarity: Normal/Reverse
Line 2: Normal/Open
Ground: Normal/Open
Test Enclosure and Earth Plus:
Patient - Applied Parts
Auxiliary (Interlead) Leakages
Applied Part Groups
Group 1:
Name: ECG
Type: Isolated
Parts: 3
AP 1 - RA
AP 2 - RL/LL
AP 3 - LA
Appendix A # Page 188