The buttons provided on Systems Settings Menu #2 are:
PAUSE AFTER
MEASUREMENT:
After each measurement of a manual leakage test, ES601Plus will pause for
the indicated time before proceeding with the next measurement of the
sequence. The available settings are "NO" (no pause); 1 second (factory
default); 2, 3, 4 and 5 seconds. The delay provided by this control allows the
user to more easily interrupt a leakage sequence in progress, so a particular
measurement of interest can be repeated as desired.
POWER UP DELAY:
During a manual leakage test, ES601Plus will wait for the indicated time
when the AC supply is applied to the DUT, before proceeding with the test
sequence. This delay allows the DUT to start up and stabilize before
subsequent leakage measurements are taken. The available options are 2
seconds (factory default), or 3 to 99 seconds in one second steps.
STOP AFTER
SUPPLY ON:
If set to "YES", this control will override the power-up delay setting
described above and will cause a manual leakage test to fu lly s to p
immediately after the AC supply is connected to the DUT by ES601Plus.
ES601Plus will then prompt the user to activate the DUT and will not
proceed until the user acknowledges by pressing the CONTINUE button.
This feature accommodates equipment with a momentary ON/OFF power
switch, which cannot be switched ON until a full AC supply is provided to
the equipment. The factory default setting for this control is "NO", which
then enables the power-up delay described above.
STOP BEFORE
SUPPLY OFF:
If set to "YES", this control will cause a manual leakage test to fu lly s to p
immediately before a test step which disconnects the AC supply from the
DUT. ES601Plus will then prompt the user to shut down the DUT and will
not proceed until the user acknowledges by pressing the CONTINUE button.
After CONTINUE is pressed, ES601Plus will then disconnect the AC supply
from the DUT. This feature accommodates equipment having special
shutdown requirements, such as diagnostic imaging equipment or equipment
incorporating a personal computer. The factory default setting for this control
is "NO".
Setup/Chapter 3 # Page 26