ES601Plus OPERATING MANUAL
CALIBRATE
OFFSET (continued):
However, OFFSET is similar to other Sequence Controls in that the OFFSET
will be displayed by ES601Plus at the start of every protective earth test,
manual or automated. This permits temporary modification of the OFFSET
value to suit a particular test lead that may be in use at the time. After the test
is completed, the OFFSET will revert to the default setting as shown on
System Settings Menu #2.
Refer to Section 5.9 for more details on operation of the protective earth test,
and for point-to-point test instructions in particular.
EXIT:
Exits to the previous menu (AUTO or MANUAL), or to AUTO MODE if
the Systems Settings menu was entered from the startup screen. On exit,
changes made to the Sequence Controls are saved in the non-volatile memory
of ES601Plus.
MORE SETTINGS:
Displays System Settings Menu #3.
If a password has been enabled via SYSTEM SETTINGS Menu #3, the user
will be prompted to enter the password when MORE SETTINGS on Menu
#2 is pressed. The password is entered via the QWERTY touch keypad.
Operations and features of the QWERTY touch keypad are discussed at the
end of this chapter.
If a password is enabled but forgotten, an override password may be obtained
from Datrend Customer Service. After the override password is entered, the
user's password can be disabled or changed.
Setup/Chapter 3 # Page 28