6.16.3 Transfer of program from Smart Card → Furnace
This function allows the transfer of programs from the Smart Card to the Multimat
®
Touch.
1. Insert the Smart Card into the slot on the right hand side of the furnace (Chip towards the
rear).
2. Touch the sensor button “M” on the numeric pad.
3. Choose the option “Smart Card”.
4. Touch the sensor button “→ Smart Card furnace” in the soft key bar.
5. Query load program confirm with “S” (Cancel with “C”).
6. Program is loaded.
7. For saving program confirm with “C”.
7. Service and maintenance
7.1 Lift
The lift is used for the vertical transportation of the firing chamber. The lift is driven by means
of a geared motor. If there is a power failure in the unit, the firing chamber can be lifted up by
hand in order to remove the work pieces. By exerting forceful pressure from the top, the firing
chamber can be moved down manually.
Attention: Do not change the tension of the tooth belt!
7.2 Firing platform complete with firing platform carrier
The firing platform is located above the firing platform carrier. The firing platform carrier is
mounted in a vertically sprung fashion to the top of the furnace subsection. The firing
platform provides heat insulation of the firing chamber, and is also used as an object carrier.
The firing platform carrier complete with O ring will seal off the firing chamber in the event of
a vacuum firing. To avoid damage of the firing platform carrier or the O-ring the firing platform
has to be placed on the platform carrier at all times.
Maintenance note: The O ring of the firing platform carrier is to be kept clean and to be
checked occasionally for possible damage.
Attention: To prevent damaging of the firing platform carrier and on the O-Ring seal place
always the firing or pressing platform onto the firing platform carrier.
7.3 Vacuum pump
The Multimat
®
Touch can be operated with all efficient vacuum pumps that have a grounded
plug connector (current consumption 2.5A max.), preferably with Dentsply vacuum pumps.
The vacuum pump should have a suction capacity of at least 30 l/min and should reach a
minimum end pressure of 30 hPa. How to connect such pumps is described in Section 3.
"Setup and initial use".
Power cord between furnace and pump must not exceed a length of 2 m.
Maintenance note: For maintenance, the instructions contained in the pump operating
manual must be complied with.
Important note: If the vacuum pump is oil-lubricated, change the oil every 3 months.