EasyManua.ls Logo

DENTSPLY MULTIMAT Touch - Firing Table for Finesse Metal Ceramics (Detrey Mode)

DENTSPLY MULTIMAT Touch
57 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
50
9.2 Firing table for Finesse metal ceramics (DeTrey Mode)
(Use honey comb)
Prog #. Program
Description
Preheat/
Low
Temp.
°C
Drying
time
min.
Preheat.
min.
Vacuum
Level
hPa
Rate of
Temp.
Rise
°C/Min.
Firing /
High
Temp.:
° C
Vacuum
Time
Min.
Firing
time
min.
311 D-FIN-Powder
Opaque
450 03:00 03:00 50 90 800 00:30 01:00
312 D-FIN-Paste
Opaque
450 05:00 03:00 50 90 790 00:06 00:30
313 D-FIN-Joint
Opaque /
Margin/margin
mod.
675 03:00 07:00 50 35 770 00:06 00:30
314 D-FIN-
Correction
Opaque
450 05:00 03:00 50 90 760 00:06 00:30
315 D-FIN-1.Main
Firing
Opaque dentine,
Dentine,
Dentine effect,
enamel,
gums
450 05:00 05:00 50 35 760 00:06 00:30
316 D-FIN-2. Main
Firing
Opaque dentine,
Dentine,
Dentine effect,
enamel,
gums
450 05:00 05:00 50 35 750 00:06 00:30
317 D-FIN-
Correction
Firing
450 05:00 05:00 50 55 730 00:06 00:30
318 D-FIN-Glaze
firing/
Stains '
450 03:00 03:00 - 70 750 - 00:06
319 D-FAC-repair
porcelain
450 05:00 05:00 50 35 810 00:06 00:30
320 D-FAC-Shade
stain porcelain
450 03:00 01:00 -. 100 730 -- 01:00
D = DeTrey Mode
FIN = Finesse metal ceramic
FAC = Finesse All Ceramic
' depending on desired grade of glaze
Caution!
Firing parameters that are not listed in the firing table should be set to "0" during
value input.
Attention: Do not use pressing platform for firing.
Note: If necessary, carry out a firing test and adjust firing temperatures or times.
Incase of firing a larger number of objects it may be necessary to adjust the firing
temperature or the holding time.