E Technical Information | Dexcom G7 User Guide 127
reading on your receiver or mobile app, these tables show you how likely it is that the
reading matches your blood glucose level (measured by YSI in the study).
For example, the total number of data pairs considered in the analysis was 47,261. Of
these, 94.6% of the sensor readings fall within ± 20 mg/dL of the YSI blood glucose
values < 70 mg/dL and within ± 20% of YSI blood glucose values 70 mg/dL.
Table 1-A. G7 accuracy to comparator method (n= 453)
Patient
Population
Number of
subjects
Total number of
paired
CGM
1
-YSI
Percent within
20/20% YSI
Percent within 20/20%
YSI on Day 1
MARD
(%)
Overall
(7 and older)
430 47,261 94.6 86.1 8.2
Adults 308 39,193 94.6 86.1 8.2
Pediatrics
(7-17)
122 8,068 94.6 86.1 8.1
Pediatrics
(2-6)*
Arm 14 148 96.6 NA 7.7
Upper
buttocks
13 143 85.3 NA 11.7
* No YSI measurements were taken for this age group; results presented are from in-
clinic CGM-SMBG matched paired measurements.
1
CGM readings are within 40-400 mg/dL, inclusive.