Table 3-C. Concurrence of G7 sensor readings and YSI values by YSI glucose
range (adults, n=308)
CGM
sensor
reading
range
(mg/dL)
YSI value range (mg/dL)
<40 40- 60 61- 80
81-
120
121-
160
161-
200
201-
250
251-
300
301-
350
351-
400
>400
<40
16
61.5%
105
4.7%
97
1.5%
3
0.0%
3
0.1%
--- --- --- --- --- ---
40- 60
9
34.6%
1,418
63.6%
944
14.8%
71
1.0%
2
0.0%
--- --- --- --- --- ---
61- 80
1
3.8%
657
29.4%
4,134
65.0%
678
9.4%
11
0.2%
4
0.1%
--- --- --- --- ---
81- 120
---
49
2.2%
1,175
18.5%
5,570
77.5%
671
12.6%
30
0.7%
7
0.2%
--- --- --- ---
121- 160
---
2
0.1%
2
0.0%
855
11.9%
3,950
74.0%
636
15.2%
58
1.5%
5
0.1%
--- --- ---
161- 200
--- ---
6
0.1%
9
0.1%
691
12.9%
2,916
69.5%
668
16.8%
55
1.4%
4
0.1%
--- ---
201- 250
--- --- --- ---
9
0.2%
608
14.5%
2,687
67.8%
651
16.8%
51
1.1%
6
0.3%
3
1.7%
251- 300
--- --- --- --- ---
4
0.1%
538
13.6%
2,604
67.3%
1,187
25.9%
77
4.4%
---
301- 350
--- --- --- --- --- ---
7
0.2%
547
14.1%
2,795
61.0%
598
33.9%
6
3.4%
351- 400
--- --- --- --- --- --- ---
9
0.2%
530
11.6%
940
53.3%
48
27.0%
>400 --- --- --- --- --- --- --- ---
14
0.3%
143
8.1%
121
68.0%
Total 26 2,231 6,358 7,186 5,337 4,198 3,965 3,871 4,581 1,764 178