E Technical Information | Dexcom G7 User Guide 135
Table 3-D. Concurrence of G7 sensor readings and YSI values by YSI glucose
range (pediatrics*; n=122)
CGM
sensor
reading
range
(mg/dL)
YSI value range (mg/dL)
<40 40- 60 61- 80
81-
120
121-
160
161-
200
201-
250
251-
300
301-
350
351-
400
>400
<40
3
100%
33
11.7%
33
2.4%
5
0.4%
--- --- --- --- --- --- ---
40- 60
---
186
66.0%
196
14.0%
15
1.1%
5
0.4%
--- --- --- --- --- ---
61- 80
---
61
21.6%
895
63.8%
120
8.6%
11
0.9%
2
0.2%
--- --- --- --- ---
81- 120
---
2
0.7%
278
19.8%
1,103
78.8%
168
13.3%
5
0.5%
--- --- --- --- ---
121- 160
--- --- ---
156
11.1%
908
71.8%
215
22.7%
16
1.9%
--- --- --- ---
161- 200
--- --- ---
1
0.1%
173
13.7%
605
63.8%
148
17.2%
10
1.6%
--- --- ---
201- 250
--- --- --- --- ---
122
12.9%
610
70.8%
169
27.7%
13
1.3%
--- ---
251- 300
--- --- --- --- --- ---
87
10.1%
344
56.4%
279
28.0%
3
0.8%
---
301- 350
--- --- --- --- --- --- ---
86
14.1%
589
59.0%
110
27.8%
1
1.9%
351- 400
--- --- --- --- --- --- ---
1
0.2%
114
11.4%
240
60.6%
21
38.9%
>400
--- --- --- --- --- --- --- ---
3
0.3%
43
10.9%
32
59.3%
Total
3 282 1,402 1,400 1,265 949 861 610 998 396 54
* Includes pediatric subjects 7-17 years of age; no YSI measurements were taken for
pediatric subjects 2-6 years of age.