Error 013 Clamp plate temperature high.
Cause:
The heater cable connection to the power supply is loose, or a heater
component malfunctioned.
Action:
Contact Dionex for assistance.
Error 014 Oven unable to reach temperature.
Cause:
Incorrect settings of the oven voltage switches on the ASE 150 rear
panel.
Action:
Set both oven voltage switches to match the voltage from the power
source at the installation site: either
110 (for voltages between 100
and 120 Vac) or
220 (for voltages between 220 and 240 Vac). For
details, see Section B.2.3
.
Cause:
The heater cable connection to the power supply is loose, or a heater
component malfunctioned.
Action:
Contact Dionex for assistance.
Error 015 Unable to rinse.
Cause:
The pump could not deliver the specified rinse volume within the
allotted time. The cell or lines may be plugged.
Action:
Run a rinse cycle (see Section 3.5). If the rinse runs without error, the
sample cell may be plugged. Replace the cell filter (see
Section 3.1.4
) and the cell frit (see Section 5.2).
Action:
If the rinse has the same error, there may be a blockage in the solvent
lines. Contact Dionex for assistance.