EasyManua.ls Logo

Dionex ASE 150 - Page 70

Dionex ASE 150
176 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ASE 150 Operator’s Manual
62 Doc. 065207-02 9/08
Error 013 Clamp plate temperature high.
Cause:
The heater cable connection to the power supply is loose, or a heater
component malfunctioned.
Action:
Contact Dionex for assistance.
Error 014 Oven unable to reach temperature.
Cause:
Incorrect settings of the oven voltage switches on the ASE 150 rear
panel.
Action:
Set both oven voltage switches to match the voltage from the power
source at the installation site: either
110 (for voltages between 100
and 120 Vac) or
220 (for voltages between 220 and 240 Vac). For
details, see Section B.2.3
.
Cause:
The heater cable connection to the power supply is loose, or a heater
component malfunctioned.
Action:
Contact Dionex for assistance.
Error 015 Unable to rinse.
Cause:
The pump could not deliver the specified rinse volume within the
allotted time. The cell or lines may be plugged.
Action:
Run a rinse cycle (see Section 3.5). If the rinse runs without error, the
sample cell may be plugged. Replace the cell filter (see
Section 3.1.4
) and the cell frit (see Section 5.2).
Action:
If the rinse has the same error, there may be a blockage in the solvent
lines. Contact Dionex for assistance.