4 Troubleshooting
Doc. 065207-02 9/08 63
Error 016 Unable to fill cell.
Cause:
An empty (or incorrectly installed) solvent reservoir.
Action:
Refill the solvent reservoir, if necessary. Check that the solvent
reservoir is correctly installed (see Section B.2.6
).
Cause:
A leaking static valve seal.
Action:
Check the collection vessel for liquid. If liquid is present, replace the
static valve seals (see Section 5.8
).
Cause:
A leaking fitting somewhere in the system.
Action:
Remove the right-side panel of the ASE 150 (see Section 5.4).
Inspect the following fittings for leaks (see Figure 4-1
): pump
fittings, pump check valves, pressure transducer fittings, pressure
relief valve fittings, static valve fittings, and solvent line fittings. Use
a 1/4-inch open-end wrench (P/N 049452) to tighten any leaking
fittings. Do not overtighten!
If tightening does not stop a leak, the tubing and fitting assembly
must be replaced. Contact Dionex for assistance.
Cause:
A leaking check valve (internally).
Action:
Replace both check valve cartridges (see Section 5.5).
Error 017 Remove cell during oven wait period.
Cause:
A sample cell was installed in the cell holder before the temperature
set point was reached.
Action:
Remove the cell from the cell holder. Never install the cell until the
STATUS field (on the STATUS screen) indicates OVEN READY and
the LED on the front panel
START button is flashing. For details, see
Section 3.2.4
.