Supplement – Infinity
®
Delta Series Software VF10.1 101
Supplement for Infinity Delta Series Software
Pulse Oximetry (Masimo RD SET SpO
2 via pod)
Description Value
Intended use The Infinity Masimo SET
®
pod and accessories are
indicated for use during both motion and non-motion
conditions and for patients who are well or poorly
perfused.
Parameter display: Saturation (%SpO
2), pulse rate, perfusion
Measurement range %SpO
2: 1 to 100%
Pulse rate: 25 to 239 /min
Perfusion: 0.02 to 20%
Measurement accuracy:
The following are for all Masimo RD SET sensors.
Saturation (%SpO
2) - During conditions with no motion:
0 to 69% not specified
70 to 100%:
Adults, pediatrics ±2
Neonates ±3
Saturation (%SpO
2) - During motion conditions:
0 to 69% not specified
70 to 100%:
Adults, pediatrics ±3
Neonates ±3
Pulse Rate (bpm) - During no motion conditions:
Adults, pediatrics, neonates ±3
Pulse Rate (bpm) - During motion conditions:
Adults, pediatrics, neonates ±5