Supplement for Infinity Delta Series Software
102 Supplement – Infinity
®
Delta Series Software VF10.1
Pulse oximetry (Nellcor Oximax SpO
2 via pod)
Parameter display Saturation (%SpO2), pulse rate
Measurement range %SpO2: 1 - 100%
Pulse rate:25 - 239 /min
Measurement accuracy, adult mode
(1, 2, 4)
:
Saturation (%SpO
2):
0 to 60% not specified
60 to 80% sensor-specific as follows:
Nellcor:
OxiMAX MAX-A, OxiMAX MAX-AL, OxiMAX MAX-P,
OxiMAX MAX-N, OxiMAX MAX-I, OxiMAX MAX-FAST ±3
Nellcor:
SC-A, MAX-R, OxiCliq A, OxiCliq P, OxiCliq N, OxiCliq I,
D-YS, DS-100A, OXI-A/N, OXI-P/I, FLEXMAX, FLEXMAX-P not specified
70 to 100% sensor-specific as follows:
Nellcor:
OxiMAX MAX-A, OxiMAX MAX-AL, OxiMAX MAX-P,
OxiMAX MAX-N, OxiMAX MAX-I, OxiMAX MAX-FAST
SC-A ±2
Nellcor:
OxiCliq A, OxiCliq P, OxiCliq N, OxiCliq I, FLEXMAX, FLEXMAX-P ±2.5
Nellcor:
D-YS, DS-100A, OXI-A/N, OXI-P/I ±3
Nellcor:
D-YS with D-YSE Ear Clip, D-YS with D-YSPD Spot Clip ±3.5
80 to 100% sensor-specific as follows:
Nellcor:
MAX-R ±3.5
Pulse Rate
(3)
: ±3 beats/min or ±3% (whichever is greater)
NOTES:
1) Since pulse oximeter measurements are statistically distributed, only about two-thirds of those
measurements can be expected to fall within ±1 A
rms of the value measured by a co-oximeter.
2) The Infinity Nellcor Oximax™ SpO2 SmartPod™ pulse oximeter with adult sensors has been
validated in human blood studies on healthy adult volunteers in induced hypoxia studies in the range
of 70-100% SpO2 against a laboratory co-oximeter and ECG monitor. This variation equals ±1 Arms
of the value measured by a co-oximeter.
3) The pulse rate accuracy has been validated on healthy adult volunteers during induced hypoxia
studies in the range of 70-100% SpO2 against a laboratory co-oximeter and ECG monitor. This
variation equals ±1 Arms of the pulse rate value measured by the ECG monitor.
4) A functional tester cannot be used to assess the accuracy of a Pulse Oximeter Probe or a Pulse
Oximeter monitor.