Supplement – Infinity
®
Delta Series Software VF10.1 107
Supplement for Infinity Delta Series Software
Output signals Graphical signal of the output voltage and current stimulation pulse:
A – Voltage (V
off) or Current (Ioff) when pulse is not delivered (OFF)
B – Voltage (V
on) or Current (Ion) when pulse is delivered (ON)
C – Pulse width (varies, dependent on setting)
1
Stimulation current accuracy is ±5% within this range. Stimulation current is disabled at
impedance values greater than 8900 Ω.
2
At a stimulation current of 30 mA, voltage and current values are approximately half (50%) of
those listed for 60 mA.
NOTE: There is no output in an open-circuited load condition.
Output voltage and current at stimulation current of 60mA
(2)
Load impedence 1.2 Ω 500 Ω 1000 Ω 6000 Ω
V
on 0.07 V 29.2 V 60.8 V 342.0 V
V
off 0V 0V 0V 0V
I
on 57.5 mA 59.8 mA 59.3 mA 57.0 mA
Ioff 0mA0mA0mA0mA