Supplement for Infinity Delta Series Software
98 Supplement – Infinity
®
Delta Series Software VF10.1
Non-Invasive Blood Pressure (NBP)
Non-Invasive Blood Pressure (NBP)
NOTE
For neonatal patients with high blood pressure and a heart rate close to 30 bpm, the monitor may not be
able to measure NBP within the maximum measurement time of 90 seconds.
Pressure measurement range
(Adult - 270 mmHg, 36.0 kPa)
Systolic NBP: 30 to 250 mmHg (4.0 to 33.3 kPa)
Mean NBP: 20 to 230 mmHg (2.7 to 30.7 kPa)
Diastolic NBP: 10 to 210 mmHg (1.3 to 28.0 kPa)
Pressure measurement range
(Pediatric - 180 mmHg, 24.0 kPa)
Systolic NBP: 30 to 170 mmHg (4.0 to 22.7 kPa)
Mean NBP: 20 to 150 mmHg (2.7 to 20.0 kPa)
Diastolic NBP: 10 to 130 mmHg (1.3 to 17.3 kPa)
Pressure measurement range
(Neonatal - 140 mmHg, 18.7 kPa)
Systolic NBP: 30 to 130 mmHg (4.0 to 17.3 kPa)
Mean NBP: 20 to 110 mmHg (2.7 to 14.7 kPa)
Diastolic NBP: 10 to 100 mmHg (1.3 to 13.3 kPa)
Connections Quick-release hose connector with single airway
Default inflation pressure Adult (270): 160 mmHg ±10 mmHg (21.3 kPa ± 1.3 kPa)
Pediatric (180): 120 mmHg ±10 mmHg (16.0 kPa ± 1.3 kPa)
Neonatal (140): 110 mmHg ±10 mmHg (14.7 kPa ± 1.3 kPa)
Inflation pressure after a valid
measurement (±10 mmHg)
Adult (270): Previous NBPSYS + 25 mmHg (+ 3.3 kPa)
Pediatric (180): Previous NBPSYS + 25 mmHg (+ 3.3 kPa)
Neonatal (140): Previous NBPSYS + 30 mmHg (+ 4.0 kPa)
Inflation pressure after an alarm Adult (270): 160 mmHg ±10 mmHg (21.3 kPa ± 1.3 kPa)
Pediatric (180): 120 mmHg ±10 mmHg (16.0 kPa ± 1.3 kPa)
Neonatal (140): 110 mmHg ±10 mmHg (14.7 kPa ± 1.3 kPa)
Maximum inflation pressure Adult (270): 265 mmHg ±5 mmHg (35.3 kPa ± 0.7 kPa)
Pediatric (180): 180 mmHg ±10 mmHg (24.0 kPa ± 1.3 kPa)
Neonatal (140): 142 mmHg ±10 mmHg (18.9 kPa ± 1.3 kPa)
Minimum inflation pressure Adult (270): 110 mmHg ±10 mmHg (14.7 kPa ± 1.3 kPa)
Pediatric (180): 90 mmHg ±10 mmHg (12.0 kPa ± 1.3 kPa)
Neonatal (140): 70 mmHg ±10 mmHg (9.3 kPa ± 1.3 kPa)
Maximum measurement time Adult (270): 2 min ±1 sec
Pediatric (180): 2 min ±1 sec
Neonatal (140): 90 sec ±1 sec
Maximum measurement time
including a retry
Adult (270): 3 min ±1 sec
Pediatric (180): 3 min ±1 sec
Neonatal (140): 90 sec ±1 sec
Software safety cut-off Adult (270): 273 ±3 mmHg (36.4 kPa ± 0.4 kPa)
Pediatric (180): 215 ±3 mmHg (28.7 kPa ± 0.4 kPa)
Neonatal (140): 153 ±3 mmHg (20.4 kPa ± 0.4 kPa)
Hardware safety cut-off Adult (270): 300 ±30 mmHg (40.0 kPa ± 4.0 kPa)
Pediatric (180): 300 ±30 mmHg (40.0 kPa ± 4.0 kPa)
Neonatal (140): 157 ±8 mmHg (20.9 kPa ± 1.1 kPa)