Supplement – Infinity
®
Delta Series Software VF10.1 99
Supplement for Infinity Delta Series Software
Continuous Non-Invasive Arterial Blood Pressure via CNAP Pod
Static cuff accuracy ±3 mmHg (± 0.4 kPa)
Calibration range Adult and Pediatric:10 to 260 mmHg ±3 mmHg (1.3 to 34.7 kPa ±
0.4 kPa)
Neonatal: 10 to 150 mmHg ±3 mmHg (1.3 to 20.0 kPa ± 0.4 kPa)
Parameter label CNAP
Systolic CNAP-S
Diastolic CNAP-D
Mean CNAP-M
NOTE: Only values for systolic, diastolic, and
mean are displayed in the p-box. Labels are not
displayed.
Units mmHg (kPa)
Measuring range
Systolic 40 to 250 mmHg (5.3 to 33.3 kPa)
Diastolic 30 to 210 mmHg (4 to 28 kPa)
Mean 35 to 230 mmHg (4.66 to 30.7 kPa)
HR range (for CNAP measurements) 20 to 200 bpm
Accuracy ± 5 mmHg (0.6 kPa)
Display resolution 1 mmHg (0.1 kPa)
Inflation pressure
Default 120 mmHg (16 kPa)
Minimum 30 mmHg (4 kPa)
Maximum 300 mmHg (40 kPa)
Over pressure limit
Pressure 300 mmHg (40 kPa) + 10%
Response time ≤ 3 s
Deflation time ≤ 15 s (to reach < 15 mmHg)