100 Instructions for use Isolette 8000 plus SW 5.n
Operation
8.7.4 Ending kangaroo mode:
1 Touch the Exit key (A) on the kangaroo screen.
The incubator exits kangaroo mode and returns to the main screen.
The incubator continues warming therapy with the settings that were active
before kangaroo mode.
2 Open the access panel and place the patient inside the incubator.
3 If skin probes are used, ensure that they are correctly connected to the device
and attached to the patient.
4 Close the device, and confirm that locking knobs are secure.
5 Depending on the operation mode, check the following settings and adjust as
necessary:
Skin temperature
Air temperature
Oxygen (optional)
Humidity
8.8 Setting up alarm limits and display units
The Setup screen allows the user to review and change display settings for some
features:
Temperature display units in Celsius or Fahrenheit degrees
Alarm limits for temperature and oxygen
Alarm limits for kangaroo mode
The following table shows alarm limit ranges and defaults:
099
40
35
30
25
20
34.8
36.8
35.0
15%
67
24
21%
0.5
0:00
A
Ranges
Air temperature upwards deviation limit
range
+1.5 °C (+2.7 °F) to +2.5 °C (4.5 °F)
Air temperature downwards deviation
limit range
- 1.C (- 2.F) to - 2.C (- 4.F)
Skin temperature deviation limit range 0.3 °C (0.5 °F) to 1.0 °C (1.8 °F)
Oxygen deviation limit range 3% to 5%