100 Instructions for use Isolette 8000 plus SW 5.n
Operation
8.7.4 Ending kangaroo mode:
1 Touch the Exit key (A) on the kangaroo screen.
The incubator exits kangaroo mode and returns to the main screen.
✓ The incubator continues warming therapy with the settings that were active
before kangaroo mode.
2 Open the access panel and place the patient inside the incubator.
3 If skin probes are used, ensure that they are correctly connected to the device
and attached to the patient.
4 Close the device, and confirm that locking knobs are secure.
5 Depending on the operation mode, check the following settings and adjust as
necessary:
– Skin temperature
– Air temperature
– Oxygen (optional)
– Humidity
8.8 Setting up alarm limits and display units
The Setup screen allows the user to review and change display settings for some
features:
– Temperature display units in Celsius or Fahrenheit degrees
– Alarm limits for temperature and oxygen
– Alarm limits for kangaroo mode
The following table shows alarm limit ranges and defaults:
099
40
35
30
25
20
34.8
36.8
35.0
15%
67
24
21%
0.5
0:00
Ranges
Air temperature upwards deviation limit
range
+1.5 °C (+2.7 °F) to +2.5 °C (4.5 °F)
Air temperature downwards deviation
limit range
- 1.5°C (- 2.7°F) to - 2.5°C (- 4.5°F)
Skin temperature deviation limit range 0.3 °C (0.5 °F) to 1.0 °C (1.8 °F)
Oxygen deviation limit range 3% to 5%