Instructions for use Isolette 8000 plus SW 5.n 47
Assembly and preparation
6 Assembly and preparation
6.1 Unpacking
WARNING
Risk of death or serious injury or damage to the device
The device could tip over.
Use 2 persons of sufficient strength to lift or move the incubator when
unpacking.
When removing the equipment from the carton, take care not to scratch or
otherwise damage unprotected surfaces.
Open the shipping carton and remove all packing material. Carefully lift and remove
the incubator from the packaging.
If the packaging or the device is damaged, contact an authorized DrägerService
representative.
6.2 Proper assembly
WARNING
Risk of death or serious injury
If the incubator is assembled incorrectly, or parts/assemblies are not reinstalled
after cleaning or maintenance, the essential performance and/or basic safety of the
device may be compromised. Noncompliance may result in death or serious injury.
Assembly and functional check must be performed by service personnel.
After the device is assembled and before it is placed into service, perform the
Functional check procedure on page 63.
The hood/shell assembly and the adjustable trolley are shipped assembled in one
carton.
6.3 Attaching accessories or optional components
To install the IV pole, utility shelf, and rail components, refer to their assembly
instructions.
WARNING
Risk of death or serious injury
The device could tip over.
Strictly adhere to the weight and placement requirements for accessories and
optional components.