Instructions for use Isolette 8000 plus SW 5.n 149
Service
Castors and movement of the device
4 Check that the product is in good condition:
All labels are complete and legible
There is no visible damage
Fuses that are accessible from the outside are in compliance with the
specified values.
All gaskets and grommets in the patient compartment are in good condition
5 Observe the instructions for use, and check that all components and
accessories required to use the product are present.
6 Check the electrical safety in accordance with the IEC 62353 standard.
7 Check the safety equipment:
Functional integrity of all acoustic alarm signals and visual symbols
8 Check temperature correlation – Temperature indicator versus incubator
temperature:
Set temperature 35 °C
Correlation ± 0.8 °C (± 1.44 °F)
Use a thermometer with an accuracy of at least ± 0.05 °C.
Do not use a mercury-in-glass thermometer.
13.5 Maintenance
There is a power entry module that has fuses on both of the mains lines. It also has
a double pole single throw (DPST) switch, which isolates the device from the mains
when it is switched to the "OFF" position.
The following table shows the preventive maintenance intervals. This table is not
related to warranty periods and no warranty is implied by the intervals listed.
Component Interval Measure Target group
Air inlet filter Every 3 months or
when necessary
Exchange Users
Mattress When necessary Exchange Users
Grommets, gaskets When necessary
1)
Exchange Users
Cuffs and iris port
sleeves
2)
Weekly or when
necessary
1)
Exchange Users
Skin temperature probes Weekly
3)
Exchange Users
Probe covers Weekly
3)
Exchange Users
O
2 sensors
2)
Yearly or when nec-
essary
Exchange Service personnel
Latch, heat shield Every 2 years Exchange Service personnel
O
2 solenoid valve
2)
Every 2 years Exchange Service personnel
Fan motor service kit Every 3 years Exchange Service personnel
Fan in sensor module Every 3 years Exchange Service personnel
O
2 pressure reducer
2)
Every 6 years Exchange Service personnel