Instructions for use Isolette 8000 plus SW 5.n 33
Overview
4.7 Auxiliary power sockets
Three auxiliary power sockets (A) are on the left side of the height adjustment
trolley.
WARNING
Risk of death or serious injury
Connecting certain electrical equipment to the auxiliary power sockets could
negatively alter the medical device resulting in a reduced level of safety, including
device malfunction.
When plugging in items other than items supplied with the system, ensure that
the total current does not exceed acceptable limits identified in the technical
data.
WARNING
Risk of reduced level of safety
Connecting other devices to auxiliary power sockets on the Isolette 8000 plus may
lead to a reduced level of safety.
Comply with the requirements of IEC 60601-1 when connecting other devices to
auxiliary power sockets on the Isolette 8000 plus.
Do not use auxiliary power sockets that are separate from the Isolette 8000 plus
CAUTION
Risk of damage to the device
The total power of all equipment connected to the auxiliary power sockets on the
trolley must be within the electrical requirements shown on the trolley.
Do not exceed the power limit shown on the trolley.
The maximum total earth leakage current of the system, including all items
plugged into the auxiliary mains outlets must not exceed 500 µA.