146 Instructions for use Isolette 8000 plus SW 5.n
Reprocessing
16 Ensure that the hand port latch with slight pressure, and open when the latch
lever is pressed.
17 If the air inlet filter is damaged, visibly dirty, or older than 3 months, replace it
(see Air inlet filter maintenance on page 151).
18 Install the air inlet filter cover, and tighten the 2 thumbscrews (see page 151).
19 On devices with the humidification system option:
A Reinstall the humidifier in the water reservoir assembly. Insert it fully into the
humidifier chamber and secure it by closing the 2 latches.
B Reinstall the lid on the water reservoir and push back the latch on top of the
lid to secure it.
C Insert the water reservoir assembly in the shell with the latch open. Then
slide it in fully and lock it in place by pushing in the latch.
20 If necessary, reattach any accessories previously removed from the device.
12.7 Preparations before reuse
1 Assemble and prepare the device so that it is ready for use, see chapter
Assembly and preparation
2 Check the operational readiness, see chapter Getting started.