Instructions for use Isolette 8000 plus SW 5.n 113
Configuration
10 Configuration
The actual screen display may differ in appearance or configuration.
10.1 System configuration menu
The system configuration menu enables the user to view and change configurable
system parameters.
10.1.1 Default settings
With power failures lasting 10 minutes or less, the set points and operation mode
are retained. With power failures lasting more than 10 minutes, the set points and
operation mode revert to factory default settings or to the settings selected in the
system configuration menu.
System configuration
menu items
Setting options Default settings
Humidity Option Yes/No No
Oxygen Option Yes/No No
Oxygen Cal Level 100%/21% 21%
Skin Control Mode Yes/No Yes
Skin Temperature Differ-
ence
Yes/No No
Language English, French, Ger-
man, Spanish, Italian,
Japanese, Dutch, Dan-
ish, Norwegian, Polish,
Portuguese, Swedish,
Finnish, Greek, Czech,
Slovak, Russian, Bulgar-
ian, Chinese, Hungar-
ian, Turkish, Lithuanian,
Serbian, Croatian, Slove-
nian, Romanian
English
Weight Units lb/kg
1)
1) For OIML devices, only kg units shall be displayed.
kg
Air Set Temp 20.0 °C to 37.0 °C (incre-
ments of 0.1 °C)
35.0 °C
Altitude
2)
2) Used for 21% oxygen calibration offset
0 ft to 12,000 ft (0 to
3657 m) (increments of
2000 ft)
0 ft (0 m)
External interface Serial Data/MEDIBUS.X Serial Data
Display Color Yellow/black / White/blue White/blue