20 Instructions for use Isolette 8000 plus SW 5.n
For your safety and that of your patients
2.7.4 Oxygen sensors
The oxygen sensor is a sealed unit that contains potassium hydroxide electrolyte.
Failure to comply with the following precautionary measures could lead to death or
serious injury.
If the sensor develops a leak, discard it immediately.
If contact with the skin or clothing occurs, rinse the area with a large quantity of
water.
In case of eye contact, flush the eye immediately for at least 15 min, holding the
eye open, and call a physician.
Use only Dräger Medical recommended oxygen sensors for proper operation.
2.7.5 Oxygen maintenance
The use of poorly maintained oxygen components increases the risk of fire and
could lead to death or serious injury.
Inspect gas/oxygen components at regular preventive maintenance intervals for
signs of corrosion or damage.
Routinely inspect oxygen cells for signs of degradation or leakage, and replace if
necessary.
2.7.6 Oxygen monitoring
Use of anesthetic agents can interfere with the accuracy of an oxygen analyzer.
Routinely monitor oxygen using an oxygen analyzer.
2.8 Use of humidity
2.8.1 Open gaps in the hood
Any open gaps in the hood reduce the incubator internal relative humidity.
Make sure that all hand port gaskets and hose grommets are properly installed
on the hood.
2.8.2 Water spillage
Water spillage may occur.
Fill the reservoir to the maximum filling limit line.
Do not overfill.
When using high levels of humidity, Dräger recommends using the optional
condensation management system.
2.9 Temperature stabilization
2.9.1 Open access panels
When the front or rear access panel is open, the temperature display may not
accurately reflect the incubator temperature.
Do not leave the access panel open longer than essential.