EasyManua.ls Logo

Dräger Isolette 8000 plus - Installing the 100% Oxygen Calibration Fixture (Optional)

Dräger Isolette 8000 plus
192 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instructions for use Isolette 8000 plus SW 5.n 57
Assembly and preparation
6.7 Installing the 100% oxygen calibration fixture (optional)
Do not install when the incubator is configured for 21% oxygen calibration (see
System configuration menu on page 113).
1 Pull down the lock on the sensor module, and slide the module out from the
hood until it stops.
2 Pull out the clip on the left side of the module, and remove the module from the
hood.
3 Rotate the locking knobs, and open the front access panel of the incubator.
4 Remove and retain the 2 mounting screws securing the cover plate slide fixture.
5 Remove the cover plate slide fixture. Also remove the sensor module slide lock.
6 Ensure the proper positioning of the O-ring (A) on the 100% calibration fixture.
The 100% calibration fixture is packaged with a small O-ring, which is coated
with a lubricant. The O-ring could possibly become loose during shipping.
008
A

Table of Contents

Other manuals for Dräger Isolette 8000 plus

Related product manuals