136 Instructions for use Isolette 8000 plus SW 5.n
Reprocessing
12.1.9 Removing the air inlet filter
To remove the air inlet filter cover located under the shell assembly on the rear of
the device, loosen the 2 thumbscrews (A).
12.2 Information on reprocessing
Instructions for reprocessing are based on internationally accepted guidelines, e.g.,
standard ISO 17664.
12.2.1 Safety information
WARNING
Risk of death or serious injury
A fire and explosion hazard exists when performing cleaning and maintenance
procedures in an oxygen-enriched environment.
► When performing cleaning and maintenance procedures, confirm that the
oxygen supply is turned off and that the equipment is disconnected from the
oxygen supply.
NOTICE
Risk of equipment damage
Some chemical cleaning agents may be conductive and leave a residue that may
permit a build-up of conductive dust or dirt.
► Do not allow cleaning agents to contact electrical components, and do not spray
cleaning solutions onto any of these surfaces.
► Use a cloth dampened with disinfectant. Do not spray cleaning solution directly
onto the surface of the device.
NOTICE
Risk of damage to the device
Harsh cleaning methods can damage equipment.
► Do not use harsh cleansers/detergents such as scouring pads or heavy-duty
grease removers or solvents, such as acetone.
046