Instructions for use Isolette 8000 plus SW 5.n 19
For your safety and that of your patients
2.6.5 High-pressure oxygen
If their ignition temperature is reached, violent ignition of oil, grease, greasy
substances, small particles or dust, dirt, or other particulate contaminants, even
small particles of metal, could occur in the presence of high-pressure oxygen.
► Take care when using high-pressure oxygen, such as that found in medical
oxygen cylinders.
2.7 Use of oxygen
2.7.1 Oxygen supply
Improper use of supplemental oxygen may be associated with serious side effects
including blindness, brain damage, and death. The risks vary with each infant.
► The qualified attending physician should prescribe the method, the
concentration, and the duration of oxygen administration.
► If it is necessary to administer oxygen in an emergency, notify the attending
physician immediately.
► Disconnect the incubator from the hospital oxygen source when oxygen is not in
use.
► Do not place auxiliary equipment that produces sparks in an incubator.
► The administration of oxygen may increase the noise level for the baby within
the incubator.
► In patient compartment, use only electrical devices that are approved for use in
an oxygen-enriched atmosphere.
► Be aware that oxygen delivered to the patient is not humidified.
2.7.2 Oxygen concentration
The oxygen concentration inspired by an infant does not accurately determine the
partial pressure of oxygen (pO
2) in the blood.
► When deemed advisable by the attending physician, measure blood pO
2 by
accepted clinical techniques.
► After each change of oxygen flow, allow at least 30 min to achieve new
concentrations.
Oxygen levels within the hood environment may be affected when the hand ports or
access panels are opened.
► Make sure that all hood hand port gaskets and hose grommets are properly
installed. Any open gaps in the hood may reduce the incubator internal oxygen.
2.7.3 Oxygen cylinders
If the gas is released rapidly due to damage or other causes, compressed gas
cylinders, such as oxygen cylinders, can become hazardous projectiles.
► Securely fasten the cylinder.
► Consult the facility gas cylinder safety procedures for the correct method to use
gas cylinders.