Instructions for use Isolette 8000 plus SW 5.n 143
Reprocessing
12.6 After reprocessing
12.6.1 Assembling the components
CAUTION
Risk of injury or damage to the device
If the incubator is assembled incorrectly, or parts/assemblies are not reinstalled
after cleaning or maintenance, the essential performance and/or basic safety of the
device may be compromised.
Assemble the device according to the instructions for use.
1 Inspect all reprocessed components for any breakage or cracks before
reassembling into the incubator.
2 On devices without the humidification system option, reinstall the manifold.
3 On devices without the condensation management system, reinstall the drain
plug in the bottom of the heater well.
4 On devices with the condensation management system, install a new collection
bottle, hose/plug assembly, and clamp (see Condensation management system
replacement (if installed) on page 152).
5 Install the heater radiator and fan impeller.
6 On devices without the humidification system option, install the heater/impeller
cover (A).
7 On devices with the humidification system option, install the heater/impeller
cover (B) and the duct cover (C).
WARNING
Risk of death or serious injury
Using older software versions can lead to incorrect operation of the humidification
system leading to injury.
The humidification system with the heater/impeller cover with duct cover must
be used with software version 5.n or higher.