94 Instructions for use Isolette 8000 plus SW 5.n
Operation
7 To confirm the Trend display selection, press the Home button (D).
Device returns to display 2.
8 Press the Keypad Lock key (E).
Keypad is locked.
8.6 Weighing– scale measurements (option)
8.6.1 Techniques for accurate scale measurements
Always weigh the infant in the center of the bed with the bed in its flat position.
Do not allow stuffed toys or other objects on the bed to lean against the
incubator walls or access panels. Inaccurate readings can occur.
Do not allow the mattress cover to touch the hood.
Secure the ventilator hose to the incubator walls in a manner that permits water
in the hose to drain away from the infant.
To ensure that they do not touch the bed, support the ventilator hose and I.V.
tubes.
Place the following items so that they return to the same relative position when
the infant has been lifted off and returned to the bed: Breathing hoses, I.V. tubes,
and sensor leads.
WARNING
Risk of death or serious injury
Failure to comply could impact infant safety.
Do not leave the infant unattended when the access panel is open.
Double-check therapeutic decisions that are based on the patient weight by
performing a reference measurement using an external scale.
WARNING
Risk of measuring error due to incorrect taring weight
If a tare weight is not determined before weighing, measuring errors may occur.
Tare the scale before each weighing process.
The bed should be level, i.e., not in the Trendelenburg or Reverse
Trendelenburg position.
8.6.2 Standard scale measurements
8.6.2.1 Initial weigh
To perform the initial weighing function:
1 If necessary, unlock the keypad.
2 At display 1, press the Display Selection.
Display 2 appears.