Instructions for use Isolette 8000 plus SW 5.n 167
Technical Data
15 Technical Data
15.1 Specifications
15.1.1 Device classification
15.1.2 Physical attributes
15.1.3 Environmental requirements
Protection class Class I, Type BF, continuous operation,
not AP
Ingress of liquids and particulate matter
(IEC 60601-1)
IPX0
Classification in accordance with EU
Directive 93/42/EEC
IIb
UMDNS code/GMDN code 12-113/36025
Height 133.3 cm to 153.7 cm (52.5 in to 60.5 in)
Width <104 cm (41 in)
Depth <76.2 cm (30 in)
Weight (without options/accessories) ≤98.5 kg (217.1 lb)
SoftBed mattress size ≥38 cm x 74 cm x 3 cm (15 in x 29.1 in x
1.2 in)
Bed Trendelenburg/reverse Trendelen-
burg tilt
Continuously variable to 12° ± 1°
Infant weight 10 kg (22 lb) maximum
Expected service life 7 years
Operating conditions
Temperature 20 °C (68 °F) to 30 °C (86 °F)
Humidity 5% to 95% relative humidity, non-con-
densing
Operating altitude Up to 3000 m (9800 ft) (per IEC60601-1)
Ambient air pressure 110 kPa to 70 kPa
Storage conditions
Temperature -20°C (-4°F) to 60 °C (140 °F)
Humidity 5% to 95% relative humidity, non-con-
densing
Ambient air pressure 110 kPa to 50 kPa