168 Instructions for use Isolette 8000 plus SW 5.n
Technical Data
15.1.4 Electrical requirements
15.1.5 General performance
Power requirements for 100/120 V~
devices
100/120 V~, 50/60 Hz, 1900 W maxi-
mum, 9.9 A maximum
Power requirements for 220/230 V~
devices
220/230 V~, 50/60 Hz, 1900 W maxi-
mum, 9.9 A maximum
Auxiliary power sockets (100 V~) (VHA
only)
100 V~, 50/60 Hz, 100 W maximum
Auxiliary power sockets (120 V~) (VHA
only)
120 V~, 50/60 Hz, 300 W maximum
Auxiliary power sockets (220/230 V~)
(VHA only)
220/230 V~, 50/60 Hz, 300 W maximum
Earth leakage current ≤ 500 µA
Air temperature mode set point range 20.0 °C (68.0 °F) to 39.0 °C (102.2 °F)
Air temperature mode set point override
temperature range
37.0 °C (98.6 °F) to 39.0 °C (102.2 °F)
Air temperature display range 15.0 °C (59 °F) to 45.0 °C (113 °F)
± 0.3 °C
Air temperature warm-up time at 22 °C
(72 °F) ambient
< 35 min
Air temperature variation < 0.5 °C (<0.9 °F)
Air temperature overshoot < 0.5 °C (<0.9 °F)
Air temperature uniformity with a level
mattress
< 0.8 °C (<1.4 °F)
Accuracy of incubator temperature indi-
cation
≤ 0.8 °C
Skin temperature mode set point range 34.0 °C (93.2 °F) to 37.0 °C (98.6 °F)
Skin temperature mode set point over-
ride temperature range
37.0 °C (98.6 °F) to 38.0 °C (100.4 °F)
Skin temperature display range 15.0 °C (59 °F) to 45.0 °C (113 °F)
± 0.3 °C, except ± 0.2 °C @ 32 °C to
42 °C
Misc specifications
Noise level within the hood environment ≤ 47 dB(A) (without servo oxygen con-
trol)
Air velocity over the mattress < 10 cm/second (4 in/second); average
of 5 points at 10 cm (4 in) above the mat-
tress
Carbon Dioxide (CO
2) level < 0.5%
Set point data retention Power failures lasting <10 min
Alarm sound level >65 dBa