Instructions for use Isolette 8000 plus SW 5.n 169
Technical Data
15.1.6 External communication
15.1.7 Humidification system (option)
COM port (output only) Only connect devices that fulfill the
requirements of the standard IEC 60950-
1 on unearthed SELV circuits or the
requirements of the standard IEC 60601-
1 on accessible secondary circuits with
max. 60 V DC nominal voltage.
Type 9-pin Sub-D (female), electrically isolated
Protocol
Configurations Serial Data Output (default) or MEDI-
BUS.X
Serial data output
Baud rate 2400
Parity None
Data bits 8
Stop bits 1
Dräger MEDIBUS.X, version 6 By the RS-232 port only
Baud rate 9600
Parity Even
Data bits 8
Stop bits 1
For a detailed description of the interface protocol, please see the manual Dräger
RS 232 MEDIBUS Protocol Definition 90 28 258 and MEDIBUS for Dräger Paediat-
ric Devices 90 29 205.
Pin assignment
Pin 2 RXD
Pin 3 TXD
Pin 5 GND
Humidity control duration of operation
after refilling
>24 hours @ 85% RH and 37 °C, in air
temperature mode
Humidity control reservoir capacity 1500 mL
Humidity control range 30% to 95% in 1% increments (at high
ambient humidity levels, low-level humid-
ity settings may not be attainable)
Humidity display accuracy between 10%
and 80% @ 20 °C (68 °F) to 40 °C
(104 °F)
± 6% RH
Humidity display range 10% to 100%
Humidity display resolution 1%
Maximum humidity levels >85% (incubator set temp at 39 °C, with
at least 30% RH at ambient)