170 Instructions for use Isolette 8000 plus SW 5.n
Technical Data
15.1.8 Oxygen control system (option)
15.1.8.1 Servo oxygen control system (option)
15.1.8.2 Manual oxygen control system (option)
15.1.9 Weighing system
15.1.9.1 Standard weighing system (option)
15.1.9.2 OIML/NAWI weighing system (option) (not available in all markets)
Oxygen inlet pressure 40 psi to 150 psi (2.8 kg/cm
2
to
10.5 kg/cm
2
)
Oxygen inlet flow rate 30 L/min
Oxygen control range 21% to 65%
Oxygen display resolution 1%
Oxygen display accuracy (100% calibra-
tion)
± 3%
Oxygen display accuracy (21% calibra-
tion)
± 5%
Oxygen display range 18% to 100%
Oxygen control accuracy ± 2% of full scale
Oxygen inlet pressure 40 psi to 150 psi (2.8 kg/cm
2
to
10.5 kg/cm
2
)
Oxygen inlet flow rate 30 L/min
Weight display range 0 kg (0 lb) to 7 kg (15.4 lb)
Weight display resolution 1 g or 1 oz
Weight display accuracy 0 to 2 kg: ±2 g (0 to 4.4 lb: ±0.07 oz)
> 2 kg: ±5 g (>4.4 lb: ±0.18 oz)
Tare weight ≤ 4.0kg (8.82lb)
Category Value
Weight display range 0 kg to 7 kg
Weight display resolution 10 g
Weight display accuracy 10 g
Maximum tare weight ≤ 4.0 kg
Verification scale interval 10 g
Scale level sensitivity 90 arc minute (1.5 degrees)
Weight display update rate 1 second